Radiometry, Photometry & Optical Calibration

Core distinctionRadiance characterizes a source and viewing direction; irradiance is incident power per area; detector counts are the end of an instrument-dependent chain. These quantities are not interchangeable.

1. Definitions and governing physics

Spectral radiance $L_\lambda$ has units W m⁻² sr⁻¹ m⁻¹, and spectral irradiance $E_\lambda$ is W m⁻² m⁻¹. For an ideal Lambertian scene under appropriate geometry, $E_\lambda=\int_\Omega L_\lambda\cos\theta\,d\Omega$.

Planck’s blackbody spectral radiance is

\[B_\lambda(T)=\frac{2hc^2}{\lambda^5}\left[\exp\left(\frac{hc}{\lambda k_BT}\right)-1\right]^{-1}.\]

Brightness temperature is the temperature of an ideal blackbody producing the measured spectral radiance at a stated wavelength; it need not equal physical temperature for a non-black surface.

Optical étendue is $G=\int n^2\cos\theta\,dA\,d\Omega$. It limits how much radiance a passive optical system can couple and is central to imaging-system throughput.

2. Detector response and uncertainty

For incident spectral power $P_\lambda$ and total throughput $T(\lambda)$, photon count expectation in integration time $t$ is

\[N_e=t\int\frac{P_\lambda T(\lambda)\lambda}{hc}\eta(\lambda)\,d\lambda,\]

where $\eta$ is detector quantum efficiency. With shot noise, dark signal $N_d$ and read noise $\sigma_r$ in electrons rms, a simplified SNR is

\[\mathrm{SNR}=\frac{N_e}{\sqrt{N_e+N_d+\sigma_r^2}},\]

assuming independent Poisson photoelectron/dark counts and Gaussian read noise. Add background, spatial nonuniformity and correlated noise for a real camera.

NEP in W/√Hz is the input optical power amplitude spectral density corresponding to unit output SNR per √Hz under a specified convention. Specific detectivity $D^*$ must state detector area and bandwidth conventions; it is not a stand-alone detector ranking across unlike measurement conditions.

3. Calibration plan for a flight camera

Calibration Question answered
Dark and bias frames How much signal exists without illumination?
Flat field How does pixel response vary spatially?
Wavelength/wavenumber calibration Where is each spectral channel centered?
Linearity and saturation How faithfully do counts track incident flux?
Absolute radiometry How do digital numbers convert to physical radiance?
Stray-light/PSF/MTF How does out-of-field or off-axis illumination contaminate the observation?
Thermal/radiation repeat Does performance drift under mission conditions?

Worked estimate: a 1 pW monochromatic signal at 1 µm corresponds to approximately $5.03\times10^6$ incident photons/s. At 60% quantum efficiency and perfect other throughput, the expected photoelectron rate is $3.02\times10^6$ electrons/s.

Modeling exercise: compute blackbody spectral radiance for 250–350 K and integrate through a realistic detector filter curve. Propagate uncertainty in filter transmission and detector responsivity separately.